Fluidics widget user interface

Fluidics widget user interface

Fluidics Control, Manual Control and Status panels

Fluidics Control, Manual Control and Status panels

Startup:

  1. Make sure fluidics system hardware is connected and powered on. Click “Initialize” to initialize hardware.

  2. Click “Load Sequences” to load the csv file you need to run the experiment. For imaging we can just include the sequences for one round.

    An example:

    sequence_name fluidic_port flow_rate volume incubation_time repeat fill_tubing_with include
    Flow SSC Buffer 25 500 2000 0 1 0 1
    Flow Probe 1 500 1500 15 1 26 1
    Flow Readout Buffer 26 500 2000 0 1 0 1
    Flow Imaging Buffer 27 500 2000 3 1 0 1
    Imaging
    Flow Stripping Buffer 28 500 2000 2 1 0 1
    Flow SSC Buffer 25 500 2000 0 1 0 1

    merfish-imaging-example.csv

  3. Introduction for control features

Priming:

  1. Fill source containers with reagents needed
  2. Remove the fitting from the syringe pump's extract port (port 2) and connect the inlet side of the flow cell connector to it
  3. In “Prime Ports”, fill in ports needed to prime, and the last port and volume you want to use for filling the system. Click “Start” and wait for it to finish.

Load sample:

  1. Mount the sample in flow cell as instructed by the flow cell manufacturer
  2. Reattach the original fitting to the syringe pump extract port and securely connect the flow cell connector
  3. Using “Manual Control”, manually flow the common buffer you used to prime the system through the flow cell with specified flow rate and volume. Fill the flow cell with this buffer, verify the connection and eliminate any bubbles.
  4. Mount the flow cell onto the stage adapter and place it on the microscope.
  5. Place the lid to block light, ensuring the tubings pass through the openings on its sides.

Imaging:

  1. Use hardware trigger and “glass slide” sample format. Find interested area in 405 nm channel with lower magnification objective. Adjust the shape/size of the area to be imaged.

    image.png

  2. Switch to higher magnification (60x) in software and save coordinates in “Wellplate Multipoint” tab

    image.png

  3. Switch to higher magnification objective (e.g., 60x silicone) on the scope and apply silicone oil